Subject: Sector 7 Management Safety Inspection Report

Division:	XFD			Location:  Sector 7 Beamline Areas
Date of Inspection:		3/17/05
Inspected by:			Efim Gluskin, Jeff Alicz and Elroy Chang

The following items were identified during the inspection as needing 
corrective action:

1.  Between 7-ID-B and 7-ID-C:  The ladders and table need to be placed 
in a more appropriate storage area (see photo below).

Ladder picture

2. Between 7-ID-B and 7-ID-C:  The electrical cabling needs to be 
better organized within this area so that the rear access doors can be 
closed on the electronic racks (see photos below).

Near 7ID-C
Near 7ID-B

3.  The chemical container outside of 7-ID-C needs a label identifying 
its contents (see photo below).

Unidentified
bottle

4.  7-ID-C:  The plastic tubing needs to be stored in a more 
appropriate manner so it doesn't fall over or block the search buttons 
(see photos below).

Plastic pipe
Steel pipes

5.  The electrical cabling in 7-ID-C needs to be better organized, and 
the ends of electrical cords need to be either taped or capped (see 
photos below)0.
cables to tidy
up in 7ID-C
Untermninated
RG45

6.  The flammable liquid storage cabinet outside of 7-ID-C needs to be 
grounded (see photo below).
Ungrounded
cabinet

7.  The electrical rack needs a back panel (see photo below).
Missing back
panel

8. The electrical cords downstream from 7-ID-D need to be better 
organized (see photos below).  Recommendation:  Once the 7-ID-D beam 
station extension is completed, the proper length of electrical cords 
should be determined in order to eliminate excess cabling throughout 
the area.

Streak camera
D cables to
tidy up