From morgan@aps.anl.gov Fri Mar 18 15:18:42 2005 Date: Fri, 18 Mar 2005 15:18:23 -0600 From: Diane MorganTo: Harold Gibson , Eric Dufresne Cc: Efim Gluskin , Edmund Chang , Jeffrey Alicz , Jin Wang , Donald Walko , Gabrielle Long , Dohn Arms , Mark Beno , Lisa Gades Subject: Sector 7 Management Safety Inspection Report Division: XFD Location: Sector 7 Beamline Areas Date of Inspection: 3/17/05 Inspected by: Efim Gluskin, Jeff Alicz and Elroy Chang The following items were identified during the inspection as needing corrective action: 1. Between 7-ID-B and 7-ID-C: The ladders and table need to be placed in a more appropriate storage area (see photo below). Ladder picture 2. Between 7-ID-B and 7-ID-C: The electrical cabling needs to be better organized within this area so that the rear access doors can be closed on the electronic racks (see photos below).
Near 7ID-C
Near 7ID-B 3. The chemical container outside of 7-ID-C needs a label identifying its contents (see photo below).
Unidentified bottle 4. 7-ID-C: The plastic tubing needs to be stored in a more appropriate manner so it doesn't fall over or block the search buttons (see photos below).
Plastic pipe
Steel pipes 5. The electrical cabling in 7-ID-C needs to be better organized, and the ends of electrical cords need to be either taped or capped (see photos below)0.
cables to tidy up in 7ID-C
Untermninated RG45 6. The flammable liquid storage cabinet outside of 7-ID-C needs to be grounded (see photo below).
Ungrounded cabinet 7. The electrical rack needs a back panel (see photo below).
Missing back panel 8. The electrical cords downstream from 7-ID-D need to be better organized (see photos below). Recommendation: Once the 7-ID-D beam station extension is completed, the proper length of electrical cords should be determined in order to eliminate excess cabling throughout the area.
Streak camera
D cables to tidy up